    Place of Origin: China
    Brand Name: Forsentek
    Certification: CE
    Model Number: FFA

    Payment & Shipping Terms:

    Minimum Order Quantity: 1pc
    Packaging Details: Shockproof carton
    Delivery Time: 3-20 working days
    Payment Terms: T/T, Western Union, MoneyGram,Paypal
    Supply Ability: 2230pcs/month
    Contact Now
    Detailed Product Description
    Material: Stainless Steel Capacity: 0-2000kg(Selectable)
    Output: Analog Force Measure Direction: Tension And Compression
    Connection: Flange-to-Flange

    Flange type compression load cell 2 ton 1ton 500kg 200kg 100kg 50kg 20kg

    Flange type compression load cell FFA is made of stainless steel utilizing strain gauge technology intended for both tension and compression force measurement. Flange to flange compression force transducer FFA offers 0-200N / 0-500N / 0-1kN / 0-2kN / 0-5kN / 0-10kN / 0-20kN seven measuring ranges to choose from,high accuracy with maximum 0.1% non-linearity of the rated capacity.Force sensor has been widely used for spring testing, welding clamp force control, hand press force verification, force distribution analysis, assembly force monitoring and other industrial applications. Besides the standard size and capacities,Forsentek is also capable of and willing to provide custom design service as per client’s request.

    Flange_type_compression_load_cell_2ton_1ton

    Dimensions(in "mm")

    compression_load_cell_2 ton_1 ton

    Ordering code Measuring range Measure direction
    FFA-200N 0~200N / 0~20kgf Tension/compression
    FFA-500N 0~500N / 0~50kgf
    FFA-1kN 0~1000N / 0~100kgf
    FFA-2kN 0~2000N / 0~200kgf
    FFA-5kN 0~5000N / 0~500kgf
    FFA-10kN 0~10kN / 0~1000kgf
    FFA-20kN 0~20kN / 0~2000kgf

    Specifications

    Capacity

    200N / 20kg / 45lb
    500N / 50kg / 112lb
    1000N / 100kg / 225lb

    2000N / 200kg / 450lb
    5000N / 500kg / 1.12klb
    10kN / 1000kg / 2.25klb

    20kN / 2000kg / 4.5klb

    Rated output 2.0 mV/V
    Excitation 3-15V
    Zero balance ±0.05mV/V
    Nonlinearity ±0.1% R.O.
    Hysteresis ±0.1% R.O.
    Nonrepeatability ±0.05% R.O.
    Creep(5min) ±0.05% R.O.
    Safe overload 150% F.S.
    Ultimate overload 200% F.S.
    Compensated temperature -10...+40oC
    Operating temperature -20...+60oC
    Temperature shift zero ±0.02% R.O./oC
    Temperature shift span ±0.02% R.O./oC
    Input resistance 400±50 ohms
    Output resistance 400±10 ohms
    Insulation resistance >5000M ohms
    Ingress protection IP62
    Material of element Stainless steel
    Cable Ø3*4000mm 4-core shielded
    Wiring code

    Red--Excitation+   Black--Excitation-

    Green--Signal+   White--Signal-

     

    Load cell output: = (Power supply * Rated output) * Load / Capacity
    Rated output is normally within 1mV/V~3mV/V.
    Power supply is normally within 3-15V.
    Load / Capacity is within 0-1.
    So the load cell output signal is normally within 0-45mV(-45mV...+45mV for bidirectional load cell).
    Example one: 2.0mV/V load cell with 10V power supply,the output range is 0~20mV (-20mV...+20mV for bidirectional transducer).
    Example two: 1.0mV/V load cell with 5V power supply,the output range is 0~5mV (-5mV...+5mV for bidirectional sensor).

    Load cell + Signal amplifier:
    Load cell normally outputs -30mV...+30mV which sometimes cannot be recognized by signal processing units directly,so signal conditioner is introduced to convert mV signal into 0-5V/0-10V/4-20mA for easy signal processing.
    Example: Thrust force load cell FFA-10kN working with 0-10V amplifier,the load-output relationship will be 0-10kN load VS 0-10V output,it is linear within the error range.

    Load cell + Digital indicator
    Digital indicator can work directly with load cell without signal amplifier,here are some functions from the indicator:
    -Display force value in real time
    -Peak holding to grab the maximum force during testing
    -RS485 or RS232 interface
    -0-5V/0-10V/4-20mA analog output
    -Switch output